Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
20
Utility Settings
•
Synchronize the timebase of two or more instruments.
Sample clock and
frequency
counter accuracy
The oscilloscope's timebase uses a built- in reference that has an accuracy
of 15 ppm. This is sufficient for most uses. However, if you are looking at
a window that is very narrow compared to the selected delay (for
example, looking at a 15 ns pulse with the delay set to 1 ms), significant
error can be introduced.
Using the built- in sample clock, the oscilloscope's hardware frequency
counter is a 5- digit counter.
See
"To supply a sample clock to the oscilloscope"
on page 340.
Supplying an
external timebase
reference
When you supply an external timebase reference, the hardware frequency
counter is automatically changed to an 8- digit counter. In this case, the
frequency counter (
[Meas] > Select > Counter
) is as accurate as the external
clock.
See
"To synchronize the timebase of two or more instruments"
on
page 341.
For more information on the hardware frequency counter, see
"Counter"
on page 257.
To supply a sample clock to the oscilloscope
1
Connect a 10 MHz square or sine wave to the BNC connector labeled
10 MHz REF
. The amplitude must be from - 5 dBm to 17 dBm
(356 mVpp to 4.48 Vpp).
2
Press
[Utility] > Options > Rear Panel > Ref Signal
.
3
Use the Entry knob and the
Ref Signal
softkey to select
10 MHz input
.
A locked padlock icon appears at the top of the display.
C A U T I O N
Maximum input voltage at 10 MHz REF connector
Do not apply more than 20 dBm Max (6.32 Vpp Max) at the 10 MHz REF BNC connector
on the rear panel or damage to the instrument may occur.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...