Vertical Controls
3
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
83
To specify channel input impedance
1
Press the desired channel key.
2
In the Channel Menu, press
Imped
(impedance); then, select either:
•
50 Ohm
— matches 50 ohm cables commonly used in making high
frequency measurements, and 50 ohm active probes.
When
50 Ohm
input impedance is selected, it is displayed with the
channel information on- screen.
When AC coupling is selected (see
"To specify channel coupling"
on
page 82) or excessive voltage is applied to the input, the oscilloscope
automatically switches to
1M Ohm
mode to prevent possible damage.
•
1M Ohm
— is for use with many passive probes and for
general- purpose measurements. The higher impedance minimizes the
loading effect of the oscilloscope on the device under test.
This impedance matching gives you the most accurate measurements
because reflections are minimized along the signal path.
See Also
•
For more information on probing, visit:
"www.agilent.com/find/scope_probes"
•
Information about selecting a probe can be found in document number
"
Agilent Oscilloscope Probes and Accessories Selection Guide
(part
number 5989- 6162EN)", available at
"www.agilent.com"
.
To specify bandwidth limiting
1
Press the desired channel key.
2
In the Channel Menu, press the
BW Limit
softkey to enable or disable
bandwidth limiting.
N O T E
When you connect an AutoProbe, self-sensing probe, or a compatible InfiniiMax probe, the
oscilloscope automatically configures the analog input channels to the correct impedance.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...