Getting Started
1
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
35
Auto Scale finds, turns on, and scales any channel with a repetitive
waveform that has a frequency of at least 25 Hz, a duty cycle greater than
0.5%, and an amplitude of at least 10 mV peak- to- peak. Any channels
where no signal is found are turned off.
The trigger source is selected by looking for the first valid waveform
starting with external trigger, then continuing with the lowest number
analog channel up to the highest number analog channel, and finally (if
digital probes are connected) the highest number digital channel.
During Auto Scale, the delay is set to 0.0 seconds, the horizontal time/div
(sweep speed) setting is a function of the input signal (about 2 periods of
the triggered signal on the screen), and the triggering mode is set to Edge.
Compensate Passive Probes
Each oscilloscope passive probe must be compensated to match the input
characteristics of the oscilloscope channel to which it is connected. A
poorly compensated probe can introduce significant measurement errors.
1
Input the Probe Comp signal (see
“Input a Waveform"
on page 32).
2
Press
[Default Setup]
to recall the default oscilloscope setup (see
“Recall
the Default Oscilloscope Setup"
on page 33).
3
Press
[Auto Scale]
to automatically configure the oscilloscope for the
Probe Comp signal (see
“Use Auto Scale"
on page 33).
4
Press the channel key to which the probe is connected (
[1]
,
[2]
, etc.).
5
In the Channel Menu, press
Probe
.
6
In the Channel Probe Menu, press
Probe Check
; then, follow the
instructions on- screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust
the trimmer capacitor on the probe for the flattest pulse possible.
On the N2862/63/90 probes, the trimmer capacitor is the yellow
adjustment on the probe tip. On other probes, the trimmer capacitor is
located on the probe BNC connector.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...