23-4
Timebase Commands
PRECision:TREFerence
PRECision:TREFerence
N O T E
The Precision Timebase feature requires the installation of the Agilent 86107A Precision Timebase Module.
Command
:TIMebase:PRECision:TREFerence
This command sets the time reference. If the time reference fails to set, an error is produced.
Query
:TIMebase:PRECision:TREFerence?
This query returns whether the time reference has been successfully set. It does not indicate
whether the time reference is still valid.
A return value of 1 indicates the time reference was successfully set the last time the :Time-
base:Precision:Treference command was sent (or the "Reset Time Reference" button was
selected).
A return value of 0 indicates the time reference was not successfully set either by the :Time-
base:Precision:TReference command or by the "Reset Time Reference" button on the front
panel. The usual causes for not being able to set the time reference is:
• the signal is not present.
• the signal is too small or too large.
• the frequency is not in the specified ranges.
This query does not indicate whether the time reference is invalid due to a change in either
frequency or amplitude of the time reference signal. Use
to query the
Precision Timebase Event Register to identify whether the timebase reference is still valid.
Returned Format
[:TIMebase:PRECision:TREFerence] {0 | 1}
Example
10 OUTPUT 707;":TIMEBASE:PRECISION:TREFERENCE?"
RANGe
Command
:TIMebase:RANGe <full_scale_range>
This command sets the full-scale horizontal time in seconds. The range value is ten times the
time-per-division value. Range is always set in units of time (seconds), not in bits.
<full_scale_range> is the full-scale horizontal time in seconds.
N O T E
In Jitter Mode, scale and position controls are disabled. Do not use this command in Jitter Mode. It generates a
“Settings conflict” error.
Query
:TIMebase:RANGe?
The query returns the current full-scale horizontal time.
Returned Format
[:TIMebase:RANGe] <full_scale_range><NL>
Examples
10 OUTPUT 707;":TIMEBASE:RANGE 10E-3"
REFerence
Command
:TIMebase:REFerence {LEFT | CENTer}
This command sets the delay reference to the left or center side of the display.
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...