7-4
Calibration Commands
CANCel
C A U T I O N
The input circuits can be damaged by electrostatic discharge (ESD). Avoid applying static discharges to the front-
panel input connectors. Momentarily short the center and outer conductors of coaxial cables
prior
to connecting
them to the front-panel inputs.
Before
touching the front-panel input connectors be sure to first touch the frame
of the instrument. Be sure the instrument is properly earth-grounded to prevent buildup of static charge. Wear a
wrist-strap or heel-strap.
CANCel
Command
:CALibrate:CANCel
During a calibration, this command is equivalent to clicking
Cancel
on a displayed calibration
message. Whenever a calibration message is displayed on the instrument, send the :CALi-
brate:CONTinue, :CALibrate:CANCel, or :CALibrate:SDONE commands. Sending any other
command, including *OPC, disrupts the instrument forcing you to cycle instrument power.
Example
10 OUTPUT 707;":CALIBRATE:CANCEL"
CONTinue
Command
:CALibrate:CONTinue
During a calibration, this command is equivalent to clicking
Continue
on a displayed calibration
message. Whenever a calibration message is displayed on the instrument, send the :CALi-
brate:CONTinue, :CALibrate:CANCel, or :CALibrate:SDONE commands. Sending any other
command, including *OPC, disrupts the instrument forcing you to cycle instrument power.
Example
10 OUTPUT 707;":CALIBRATE:CONTINUE"
ERATio:DLEVel?
Query
:CALibrate:ERATio:DLEVel? CHANnel<N>
This query returns the last dark level value for the specified channel, regardless of the cur-
rent calibration status. If an extinction ratio calibration has been performed the returned
value is the calibration result. If no calibration has been performed the default value of 0.0 is
returned. <N> is an integer, from 1 to 4.
Returned Format
[:CALibrate:ERATio:DLEVel] <value><NL>
ERATio:STARt
Command
:CALibrate:ERATio:STARt CHANnel<N>
Starts an extinction ratio calibration. Before performing an extinction ratio calibration, dis-
play an eye diagram and adjust the vertical scale and offset so that the eye diagram uses the
full display. Also, the dark level (the signal level when there is no input to the measurement)
must be on the screen to be correctly measured.
Whenever a calibration message is displayed on the instrument, send the :CALibrate:CON-
Tinue, :CALibrate:CANCel, or :CALibrate:SDONE commands. Sending any other command,
including *OPC, disrupts the instrument forcing you to cycle instrument power.
<N> is an integer, from 1 to 4.
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...