17-13
Mask Test Commands
SWAVeform
N O T E
Compatibility with the Agilent 83480A/54750A. The :MTESt:TEST ON command serves the same function and
has been retained for compatibility with the Agilent 83480A/54750A. All new programs should use the :STARt
command.
SWAVeform
Command
:MTESt:SWAVeform <source>, <destination>[,<filename>[, <format>]]
This command saves waveforms from a channel, function, or waveform memory in the event
of a failure detected by the limit test. Each waveform source can be individually specified,
allowing multiple channels,or functions to be saved to disk or waveform memories. Setting a
particular source to OFF removes any waveform save action from that source.
N O T E
This command operates on waveform and color grade gray scale data which is not compatible with Jitter Mode.
Do not use this command in Jitter Mode. It generates a “Settings conflict” error.
<source>
{CHANnel<N> | FUNCtion<N> | WMEMory<N>}
<destination>
{OFF | WMEMory<N>| DISK}
<filename>
An ASCII string enclosed in quotation marks. If no filename is specified, the assigned file-
name will be
MaskLimitChN_X
,
MaskLimitFnN_X
,
MaskLimitRspN_X
, or
MaskLimitMemN_X
, where X is an incremental number assigned by the instrument. If no
path is specified, the default path will be
D:\User Files\waveforms
. (C drive on
86100A/B instruments.)
N O T E
If the selected waveforms of consecutive limit tests are to be stored in individual files, omit the <filename>
parameter. The waveforms will be stored in the default format (INTERNAL) using the default naming scheme.
<format>
{TEXT [,YVALues | VERBose] | INTernal}
where INTernal is the default value, and VERBose is the default value for TEXT.
Example
The following example turns off the saving of waveforms from channel 1 in the event of a limit
test failure.
10 OUTPUT 707;”:MTEST:SWAVEFORM CHAN1,OFF”
Query
:MTESt:SWAVeform? <source>
The query returns the current state of the :MTESt:SWAVeform command.
Returned Format
[:MTESt:SWAVeform] <source>, <destination>, [<filename>[,<format>]]<NL>
Example
The following example returns the current parameters for saving waveforms in the event of a
limit test failure.
10 DIM SWAV$[50]
20 OUTPUT 707;”:MTEST:SWAVEFORM? CHANNEL1”
30 ENTER 707;SWAV$
SWAVeform:RESet
Command
:MTESt:SWAVeform:RESet
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...