18-21
Measure Commands
DEFine
N O T E
Using "mV" or "V" following the numeric value for the voltage value will cause Error 138-Suffix not allowed.
Instead, use the convention for the suffix multiplier as described in
Example
10 OUTPUT 707;":MEASURE:DEFINE? THRESHOLDS"
Table 18-2. :MEASure:DEFine Interactions
MEASure Commands
THResholds
TOPBase
EWINdow
CGRAde
DELTatime
TREFerence
RISEtime
x
x
x
FALLtime
x
x
x
PERiod
x
x
FREQuency
x
x
VTOP
x
VBASe
x
VAMPlitude
x
PWIDth
x
x
NWIDth
x
x
OVERshoot
x
x
DUTYcycle
x
x
DELTatime
x
x
VRMS
x
x
PREShoot
x
x
VLOWer
x
x
VMIDdle
x
x
VUPPer
x
x
VAVerage
x
x
VARea
x
x
DELTatime
x
x
x
CGRade:CRATio
x
x
CGRade:CROSsing
x
x
CGRade:DCDistortion
x
x
x
CGRade:DCYCle
x
x
x
CGRade:ERATio
x
CGRade:EHEight
x
CGRade:ESN
x
CGRade:OFACtor
x
CGRade:OLEVel
x
CGRade:PWIDth
x
x
CGRade:ZLEVel
x
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...