18-13
Measure Commands
CGRade:COMPlete
Example
The following example measures the bit rate of the displayed eye.
10 OUTPUT 707;”:MEASURE:CGRADE:BITRATE”
CGRade:COMPlete
Command
:MEASure:CGRade:COMPlete <comp_hits>
Sets the color grade measurement completion criterion. The data for color grade display is
the same as for gray scale display. Auto skew (
) also uses the current color grade
measurement completion criterion. If auto skew fails to make the bit rate measurement or
determine the time of the crossing points needed to compute the skew, it may be necessary
to increase the color grade completion criterion. Increasing the value will increase the time
for auto skew to complete, allowing it to collect more data points before executing the bit
rate and crossing time measurements. <comp_hits> is the number of hits that the peak-num-
bers-of-hits, in the color grade database, must equal or exceed before a color grade measure-
ment is executed.
Mode
Eye or Oscilloscope modes
Query
:MEASure:CGRade:COMPlete?
The query returns the current setting for color grade completion.
Returned Format
[:MEASure:CGRade:COMPlete] <comp_hits><NL>
A color grade measurement query will return 9.37 until the measurement is com-
plete.
Examples
The following example sets the completion criterion to 10 hits.
10 OUTPUT 707;”:MEASURE:CGRADE:COMPLETE 10”
The following example sets the color grade complete value, then starts a Vmax measurement
with the color grade database as the source.
10 OUTPUT 707;”:MEASURE:CGRADE:COMPLETE? 8”
20 OUTPUT 707;”:DEFINE:CGRADE ON”
30 OUTPUT 707;”:MEASURE:VMAX CGRADE”
CGRade:CRATio
Command
:MEASure:CGRade:CRATio <format> [,{CHANnel<N> | FUNCtion<N> | CGMemory}]
Measures the contrast ratio of the RZ (Return-to-Zero) eye diagram on the color graded dis-
play. The dark level or dc offset of the input channel must have been previously calibrated.
See
to perform a dark level calibration. If the source is not set,
the lowest numbered signal that is on will be the source of the measurements. <format> is
{RATio | DECibel | PERCent}.
Mode
Eye mode only. Ensure that the eye type is set to RZ. See
.
Query
:MEASure:CGRade:CRATio? <format> [,{CHANnel<N> | FUNCtion<N> | CGMemory}]
This query returns the contrast ratio of the color graded display.
Returned Format
[:MEASure:CGRade:CRATio] <value>[,<result_state>]<NL>
<value> is the contrast ratio. If SENDvalid is ON, the <result_state> is returned with the
measurement result. Refer to
for a list of the result states.
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...