18-16
Measure Commands
CGRade:ESN
percentage value that is used to offset the measured extinction ratio value. Correction factor
is always specified as a percentage, regardless of the format or units specified for extinction
ratio measurement results.
Mode
Eye mode only.
Restrictions
Software revision A.04.00 and above (86100C instruments)
Query
:MEASure:CGRade:ERFactor? CHANnel<N>
This query returns the extinction ratio correction settings for the specified channel. A correc-
tion factor value is returned regardless of whether correction is on or off.
Returned Format
[:MEASure:CGRade:ERFactor] {ON|OFF}<NL>
Example
10 OUTPUT 707; ":MEASure:CGRade:ERFactor CHANnel4,ON,80"
CGRade:ESN
Command
:MEASure:CGRade:ESN [{CHANnel<N> | FUNCtion<N> | CGMemory}]
Measures the eye signal-to-noise. The data for color grade display is the same as for gray
scale display. If the source is not set, the lowest numbered signal display that is on will be the
source of the measurement. This measurement was called Q-factor in the 83480A/54750A.
Mode
Eye mode only.
Query
:MEASure:CGRade:ESN? [{CHANnel<N> | FUNCtion<N> | CGMemory}]
The query returns the eye signal-to-noise of the color grade display.
Returned Format
[:MEASure:CGRade:ESN] <value>[,<result_state>]<NL>
<value> is the eye signal-to-noise value. If SENDvalid is ON, the <result_state> is returned
with the measurement result. Refer to
for a list of the result states.
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASURE:CGRADE:ESN?”
CGRade:EWIDth
Command
:MEASure:CGRade:EWIDth <format> [,{CHANnel<N> | FUNCtion<N> | CGMemory}]
Measures the eye width on the eye diagram of the current color grade display. The data for
color grade display is the same as for gray scale display. If the source is not set, the lowest
numbered signal display that is on will be the source of the measurement. <format> is {RATio
| TIME}. The default format is TIME.
Mode
Eye mode only.
Query
:MEASure:CGRade:EWIDth? <format> [,{CHANnel<N> | FUNCtion<N> | CGMemory}]
The query returns the eye width of the color grade display.
Returned Format
[:MEASure:CGRade:EWIDth] <value>[,<result_state>] <NL>
<value> is the eye width. If SENDvalid is ON, the <result_state> is returned with the mea-
surement result. Refer to
for a list of the result states.
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASURE:CGRADE:EWIDTH?”
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...