18-42
Measure Commands
SINTegrity:SIGNal
SINTegrity:SIGNal
Command
MEASure:SINTegrity:SIGNal {CLOCk | DATA}
This command specifies the type of signal being measured in Jitter and Signal Integrity mode.
It replaces the obsolete command :MEASure:JITTer:SIGNal.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Query
MEASure:SINTegrity:SIGNal?
Returned Format
[:MEASure:SINTegrity:SIGNal] {CLOCk | DATA}<NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:SIGNAL CLOCK”
SINTegrity:SIGNal:AUTodetect
Command
MEASure:SINTegrity:SIGNal:AUTodetect {ON | 1 | OFF | 0}
This command turns automatic detection of the signal type (clock or data) on or off. The
automatic detection happens during an autoscale. This command replaces the obsolete com-
mand :MEASure:JITTer:SIGNal:AUTodetect.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Query
MEASure:SINTegrity:SIGNal:AUTodetect?
Returned Format
[:MEASure:SINTegrity:SIGNal:AUTodetect] {1 | 0}<NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:SIGNAL:AUTODETECT ON”
SOURce
Command
:MEASure:SOURce <source>[,<source>]
Selects the source for measurements. You can specify one or two sources with this command.
All measurements except MEASure: DEFine:DELTatime are made on the first specified
source. The delta time measurement uses two sources if two are specified. If only one source
is specified, the delta time measurement uses that source for both of its parameters. The
source is always color grade/gray scale data in eye mode, except for average optical power
and histogram measurements. This is a global definition. It is used for all subsequent remote
measurements unless a different source is specified with the optional source parameter in the
measure command.
<source> is {CHANnel<N> | FUNCtion<N> | RESPonse<N> | WMEMory<N>}. <N>, for chan-
nels, functions, TDR responses and waveform memories, is 1, 2, 3, or 4.
Mode
Oscilloscope and TDR modes. Eye mode uses this for average optical power measurements.
Query
:MEASure:SOURce?
The query returns the current source selection.
Returned Format
[:MEASure:SOURce] <source>[,<source>]<NL>
Example
10 OUTPUT 707;":MEASURE:SOURCE CHANNEL1"
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...