18-44
Measure Commands
TDR:MAX
TDR:MAX
Command
:MEASure:TDR:MAX {CHANnel<N> | RESPonse<N>}
Measures the maximum TDR impedance for the selected channel or response. When used as
a query, the returned value uses the same units as the setting for the selected channel or
response. For example, if the channel units are set to volts, this query returns a value in volts.
Restrictions
TDR mode. Software revision A.05.00 and above (86100C instruments).
Query
:MEASure:TDR:MAX? {CHANnel<N> | RESPonse<N>}
Returns the maximum impedance (Y-axis ) value that is to the right side of the reference
plane.
Returned Format
[:MEASure:TDR:MAX {CHANnel<N> | RESPonse<N>}] <value><NL>
Example
10 OUTPUT 707;”:MEASure:TDR:MAX RESPONSE1”
TDR:MIN
Command
:MEASure:TDR:MIN {CHANnel<N> | RESPonse<N>}
Measures the minimum TDR impedance for the selected channel or response. When used as a
query, the returned value uses the same units as the setting for the selected channel or
response. For example, if the channel units are set to volts, this query returns a value in volts.
Restrictions
TDR mode. Software revision A.05.00 and above (86100C instruments).
Query
:MEASure:TDR:MIN? {CHANnel<N> | RESPonse<N>}
Returns the minimum impedance (Y-axis ) value that is to the right side of the reference
plane.
Returned Format
[:MEASure:TDR:MIN {CHANnel<N> | RESPonse<N>}] <value><NL>
Example
10 OUTPUT 707;”:MEASure:TDR:MIN RESPONSE1”
TMAX
Command
:MEASure:TMAX [<source>]
Measures the first time at which the first maximum voltage of the source waveform occurred.
The source is specified with the MEASure:SOURce command or with the optional parameter
following the TMAX command. In TDR mode, the time reported is measured with respect to
the reference plane. <source> is {CHANnel<N> | FUNCtion<N> | WMEMory<N> |
RESPonse<N>}. <N> is an integer, from 1 through 4.
Mode
Oscilloscope and TDR modes.
Query
:MEASure:TMAX? [<source>]
The query returns the time at which the first maximum voltage occurred.
Returned Format
[:MEASure:TMAX] <time>[,<result_state>]<NL>
<time> is the time at which the first maximum voltage occurred. If SENDvalid is ON, the
<result_state> is returned with the measurement result. Refer to
for a list of the result states. When receiving numeric data into numeric variables, turn off the
headers. Otherwise, the headers may cause misinterpretation of returned data.
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...