10-6
Disk Commands
PWAVeform:RANGe:STARt
PWAVeform:RANGe:STARt
Restrictions
Software revision 4.10 and above on an 86100C. Option 201, Advanced Waveform Analysis
Software installed.
Command
:DISK:PWAVeform:RANGe:STARt <bit_number>
Sets or queries the start bit setting for saving a range of pattern waveform bits using the
DISK:
PWAVeform
:SAVE command. <bit_number> is an integer. You must first specify that a
range of the pattern will be saved by using the DISK:
PWAVeform
:RANGe command.
Query
:DISK:PWAVeform:RANGe:STARt?
Returned Format
[:DISK:PWAVeform:RANGe:STARt] <bit_number><NL>
Example
10 OUTPUT 707;":DISK:PWAVEFORM:RANGE:START 10"
PWAVeform:RANGe:STOP
Restrictions
Software revision 4.10 and above on an 86100C. Option 201, Advanced Waveform Analysis
Software installed.
Command
:DISK:PWAVeform:RANGe:STOP <bit_number>
Sets or queries the stop bit setting for saving a range of pattern waveform bits using the
DISK:
PWAVeform
:SAVE command. <bit_number> is an integer. You must first specify that a
range of the pattern will be saved by using the DISK:
PWAVeform
:RANGe command.
Query
:DISK:PWAVeform:RANGe:STOP?
Returned Format
[:DISK:PWAVeform:RANGe:STOP] <bit_number><NL>
Example
10 OUTPUT 707;":DISK:PWAVEFORM:RANGE:STOP 20"
PWAVeform:SAVE
Command
:DISK:PWAVeform:SAVE <file_name>
Saves a pattern waveform to a file with the file extension .csv. <file_name> is the name of the
file, with a maximum of 254 characters (including the path name, if used). The file name
assumes the present working directory if a path does not precede the file name. The data is
saved in an ASCII comma separated file (csv), with the amplitude data for each source (chan-
nel or function) placed in a separate column. In addition to amplitude values, saved pattern
waveform files include a header of setup information.
Patterns that include a large number of bits and high resolution involve large amounts of
data. Saving these files may require several hours and one or two gigabytes (GB) of memory.
Use *OPC or *OPC? with this command in order to synchronize data acquisition with remote
control.
If you plan on loading a saved pattern waveform back into the instrument, be sure to also save
the instrument setup. You will need to load (restore) the instrument settings at the same
time that you load the associated pattern waveform.
Restrictions
Software revision 4.10 and above on an 86100C. Option 201, Advanced Waveform Analysis
Software installed. Eye/Mask or Oscilloscope instrument mode with pattern lock triggering.
One or more channels or functions (invert, subtract, or magnify) turned on. Optional MAT-
LAB Filter and Linear Feedforward Equalizer applications closed (not running).
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...