18-31
Measure Commands
JITTer:LEVel?
Restrictions
Jitter mode. Software revision A.04.00 and above (86100C instruments).
Returned Format
[:MEASure:JITTer:ISI] <value><NL>
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASure:JITTer:ISI?”
JITTer:LEVel?
Query
:MEASure:JITTer:LEVel?
Returns the amplitude level at which jitter measurements are made.
Restrictions
Jitter mode. Software revision A.04.00 and above (86100C instruments).
Returned Format
[:MEASure:JITTer:LEVel] <value><NL>
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASure:JITTer:LEVel?”
JITTer:LEVel:DEFine
Command
:MEASure:JITTer:LEVel:DEFine {PERCent,<percentage_value> | UNITs,<level_value> | AVERage}
Defines the jitter sampling level. It may be specified as a percentage in the range of 30% to
70%, as an absolute amplitude level, or as the average amplitude of the test signal. If you
specify UNITs, the level value is interpreted as Watts or Volts depending on the type of input
channel selected: optical or electrical. For example, if a value of 5.00E-3 is entered, it will be
interpreted as 5 mW when applied to an optical channel and 5 mV when applied to an electri-
cal channel.
Restrictions
Jitter mode. Software revision A.04.00 and above (86100C instruments).
Query
:MEASure:JITTer:LEVel:DEFine?
This query returns the current setting for the jitter sampling level.
Returned Format
[:MEASure:JITTer:LEVel:DEFine] <value><NL>
Example
:MEASure:JITTer:LEVel:DEFine PERCent,40
JITTer:PATTern?
Query
:MEASure:JITTer:PATTern?
Returns definite-length block data. The data block contains the pattern as determined by the
instrument. Each value in the pattern is a single byte. Values in the pattern are the ASCII
values for '0' and '1' (30 hex and 31 hex, respectively). The data block is followed by a termi-
nator character, 0A hex (linefeed). This query will return an incomplete description of the
pattern if all of the data needed to determine the pattern has not yet been acquired. This
query produces an error if jitter signal type is set to clock signal.
Use the
:MEASure:JITTer:DDJVsbit? query to return the
DDJ values. Use the
:MEASure:JITTer:EBITs?
query to return the
bit numbers.
Restrictions
Jitter mode. Software revision A.04.00 and above (86100C instruments). When writing new
code for software revision A.07.00 and above, use the recommended command
Returned Format
[:MEASure:JITTer:PATTern] <value><NL>
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...