7.6
Manual cleaning with wipe disinfection
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*
Validated with enzyme cleaner “Cidezyme Johnson & Johnson”
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
►
Mobilize non-rigid components during cleaning.
►
Clean the product under running water, using a suitable cleaning brush
until all visible residues have been removed from the surfaces.
►
Brush cannulation with cleaning brush TA011944 and difficult to
access surfaces with a suitable cleaning brush made of plastic for at
least 1 min.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available on Aesculap Extranet at
https://extranet.bbraun.com).
Phase II
►
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
►
Spray products with a pH neutral enzyme solution, let soak in for at
least 2 minutes and then wipe off.
Phase III
►
Mobilize non-rigid components during cleaning.
►
Rinse product under running tap water for at least 5 minutes.
►
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
►
Contamination should be removed with a lint-free cloth or soft brush
moistened with enzyme cleaner.
►
Flexible components (such as sleeves) and cannulations should be
rinsed for 20 seconds with the water pistol (cold water, at least
2.5 bar).
►
After manual cleaning, check visible surfaces and areas of flexible com-
ponents for residues.
►
If necessary, repeat the cleaning process (phase 1 to 3).
Phase IV
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase V
►
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase VI
►
Rinse disinfected surfaces after the prescribed reaction time for at least
1 minute under running demineralized water.
►
Drain any remaining water fully.
Stage VII
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemicals
I
Pre-cleaning
RT (cold)
≥
2
-
D-W
Until visually clean
II
Cleaning with enzyme solution
RT (cold)
≥
2
0.8
D–W
pH-neutral*
III
Intermediate rinse
RT
≥
5
-
D-W
–
IV
Drying
RT
-
-
-
-
V
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
VI
Final rinse
RT (cold)
0.5
-
FD–W
-
VII
Drying
RT
-
-
-
-