6247G/6247C DC Voltage Current Source/Monitor Operation Manual
9.2 Source and Measurement Function
9-9
Voltage measurement input leak current:
100 pA or lower
GPIB (6247G only):
Compliant with IEEE-488.2-1987
Interface function;SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0,
E2
Connector;Amphenol 24 pin
USB:
USB 2.0 Full-speed
Connector; Type B
RS-232 (6247C only):
Compliant with EIA232
Baud rate; 19200, 9600, 4800, 2400, 1200, 600, 300
Parity; Even, Odd and None
Number of data bits; 7 bits, 8 bits
Number of stop bits; 1 bit, 2 bits
Connector; Dsub 9 pin
External control signal:
TRIGGER IN,
INTERLOCK, OPERATE IN/OUT,
OPERATE OUT, SYNC OUT
Connector; BNC