6247G/6247C DC Voltage Current Source/Monitor Operation Manual
9.3 Setting Time
9-10
9.3
Setting Time
Minimum pulse width:
50
s
Minimum step (repeat) time:
Under fixed source/measurement range, integration time of 100
s, the
minimum measurement or source delay time, calculation function
OFF, and voltage/current measurement
Source delay time:
Period (pulse cycle):
Pulse width:
Measurement
Memory mode
Minimum step time
OFF
- 0.5
ms
BURST
2 ms
ON
NORMAL
10 ms
OFF
Setting range
Resolution
*6
Setting accuracy
0.030 ms to 60.000 ms
1
s
60.01 ms to 600.00 ms
10
s
(0.1% + 10
s)
600.1 ms to 6000.0 ms
100
s
6001 ms to 59998 ms
1 ms
Setting range
Resolution
*6
Setting accuracy
0.500 ms to 60.000 ms
1
s
60.01 ms to 600.00 ms
10
s
(0.1% + 10
s)
600.1 ms to 6000.0 ms
100
s
6001 ms to 60000 ms
1 ms
Setting range
Resolution
*6
Setting accuracy
0.050 ms to 60.000 ms
1
s
60.01 ms to 600.00 ms
10
s
(0.1% + 10
s)
600.1 ms to 6000.0 ms
100
s
6001 ms to 59998 ms
1 ms