6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.12 Measurement Data Memory Function
5-65
NOTE:
In the following cases, store mode ON/OFF and storing operation changes cannot be performed.
• During auto trigger mode in the DC or pulse source mode
• During sweep operation in the sweep source mode
5.2.12.2
Memory Clear
The memory data is cleared in the following conditions:
•
When Memory Clear is executed on the menu screen
•
When the RL command is executed
•
When the store mode is switched from OFF to Normal
•
When the store mode is switched from Burst to Normal
•
Power ON
Table 5-16 Comparison of Memory Store Operations
Normal
Burst
Recommended usage
Low-speed measurement
When storing measured data for regular
measurement such as DC or pulse mea-
surement
High-speed measurement
When reading the measured data after
measuring a certain number of times such
as sweep measurement
Minimum repeat time (*)
10 ms
2 ms
Measurement data display
Displayed in real time
Displayed in idle time of measurement
tasks, or when measurement is not per-
formed.
Data output
Reads the latest data by the
ENTER key.
Available
Not available
RECALL and RN1 commands
Available
Operation at memory full
The ST indicator blinks.
MFL (bit 10) of the device event status register is set.
Storing data is stopped.
Measurement stops.
Sweep mode: STOP
DC or pulse mode: HOLD
Comparator
calculation
results
Complete Out HI/GO/LO signal
Output in real time
Not output
Buzzer
HI/LO/GO display
Displays in idle time of measurement
tasks, or when measurement is not per-
formed.
Flag for the specified number of memory stored
data reached
This flag bit is set when the number of measured data reaches the specified number
of stored data in the measurement buffer memory (using the RNM command).
Device Event Status Register ASM (bit4)
(*) Integration time: 100
s. Source delay: 30
s. Measurement delay: 100