5.1.5 Connection with Fixture 12701A
5-7
5.1.5
Connection with Fixture 12701A
The following figure shows the connection with the 12701A.
4-wire connection is used.
2-wire connection does not require the SENSE connection.
For more information on the device connection inside the 12701A, refer to 12701A Operation Manual.
Figure 5-7 Connection with 12701A
CAUTION: Follow the procedure below to prevent electric shock.
1.
Be sure to ground the 12701A protective grounding terminal
.
2.
Connect the 12701A LID SIGNAL to the INTERLOCK terminal on the 6247G/6247C rear panel, and set the
parameter "OPR Signal" to InterLock In.
This enables the interlock function. When the 12701A cover opens, the 6247G/6247C will be set to Standby.
To INTERLOCK input terminal
on the rear panel
6247G/C