6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.3 Operations in Sweep Source Mode
5-14
2. Changing the sweep measurement parameters
The sweep measurement parameters are basically changeable only in Standby or Suspend status, but
the following parameters are changeable during sweep stop in Operate status.
•
Time parameter
• Hold time
• Source delay time
• Measurement delay time
• Pulse width
• Period time
•
Start address and stop address for random sweep
(These are changeable in the start address to stop address range when switching from Standby
status to Operate or Suspend.)
•
Sweep function parameter
• Repeat count
• Reverse mode ON or OFF
• RTB ON or OFF
• Measurement auto range ON or OFF
• Measurement ON or OFF
• Measurement integration time
• Selecting COMPLETE OUT or SYNC OUT external control signal output function
3. Indicator display for sweep status
indicates the sweep status.
Sweeping:
The indicator rotates.
HOLD:
The Indicator stops rotation but remains displayed.
STOP:
Indicator is not displayed.