6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.3 Operations in Sweep Source Mode
5-13
1. Setting the sweep tyme
•
Select
DC-SWP
for DC Sweep and
PLS-SWP
for pulse sweep by using the
MODE
key.
•
Select the item
B) SWEEP
by using the
MENU
key.
Select
Linear
for linear Sweep,
Fixed
for fixed sweep,
Random
for random sweep or
Linear2
for 2-slope linear sweep from
1) Sweep Type
.
NOTE:
The maximum number of sweep steps is 8000. If the number of steps is set to more than this, the
error message "801 Over Step" is displayed in Operate status.
Pulse
sweep
Linear sweep
Sweeps with staircase pulse wave-
forms of the step value between the
first value and the last value.
Fixed sweep
Sweeps with pulse waveforms of the
specified constant value the specified
number of sample count.
Random sweep
Sweeps with pulse waveforms of the
source values stored in the memory
from the start address to the stop
address.
2-slope linear sweep
Sweeps with staircase pulse wave-
forms of the 1st step value between
the first value and the middle value,
and sweeps with staircase pulse wave-
forms of the 2nd step value between
the middle value and the last value.
Table 5-4 Operations in Sweep Source Mode (2 of 2)
Sweep types
Description
Waveform