6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.6 Pulse Measurement
2-40
Preparation
1. Follow the same procedure described in Section 2.2.5, "DC Measurement."
Setting the pulse source value
2. Press
123…
,
2
, and
ENTER
in order.
The pulse source value is set to 2 V.
Setting the current limit value
3. Set
Balance using
MENU
,
SOURCE
and
LMT Input
.
4. Press
LIMIT
,
123...
,
3
and
ENTER
in order.
The current limit value is set to
3 mA.
5. Press
LIMIT
.
The HOME screen is displayed.
Setting the base value
6. Press
MENU
.
7. Press
or
to select
A) SOURCE
. Press
to enter the Select level.
8. After confirming the item is
1) PLS Base
, press
to enter the Input/Run level,
and press
123…
,
1
and
ENTER
in order.
The base value is set to 1 V.
Setting the pulse time
9. Press
,
,
,
,
, and
in this order.
The time setting select level
D) TIME
is displayed.
10. Press
to
select
2) Src Delay
. Press
to enter the Input/Run level.
11. Press
123..., 1,
and
ENTER
in order.
Tds is set to 1 ms.