6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.5 DC Measurement
2-38
11. Return the source value to +2 V with
or
, and press
123...,
6
and
ENTER
in order.
The voltage source value is set to 6 V in the 50 V range.
As the source current is limited by the current limiter, the limiter indicator goes
ON. (See Point D in Figure 2-15.)
Current source (ISVM)
12. Press
VS/IS
.
The measurement changes to current source and the 6247G/6247C sets Suspend
status.
13. Press
123..., 2, UNIT
,
ENTER
and
OPR
in order.
Cursor (blinking)
Blinking
Cursor (blinking)