6247G/6247C DC Voltage Current Source/Monitor Operation Manual
4.2.4 HOLD Key (Trigger Mode)
4-4
4.2.4
HOLD Key (Trigger Mode)
Switches the source and measurement trigger mode.
4.2.5
LIMIT Key (Limiter Setting)
Switches between the source value setting screen and the limit value setting screen.
For more information on the limit value setting, refer to Section 2.2.2, "Setting Limiter Values."
4.2.6
MENU Key (Parameter Setting)
Pressing the
MENU
key displays the parameter group setting screen.
Select an item from
(
A) SOURCE
to
M) SYSTEM
)
in the Category level menu with
or
.
For more information on set items and the setting method, refer to Section 2.2.3, "Menu Operation."
SOURCE
Sets source-related common elements.
PLS Base
Sets the base value at pulse source.
Suspend V
Sets the output voltage in Suspend.
Suspend Z
Sets the output impedance in Suspend.
HiZ:
The output current limit is set to
100 nA, so the output
impedance increases.
LoZ:
The output current limit is set to the current limit value
at voltage source or 30 digits in the current range (100
digits in the 3
A range) at current source, so the output
impedance decreases.
Source mode
AUTO
HOLD
DC/Pulse
Repeats source and measurement within
the period time of the time parameter.
The sampling indicator turns ON during
measurement.
Executes source and measure-
ment with trigger input.
The HOLD indicator turns ON.
Sweep
Repeats source and measurement within
the period time of the time parameter.
The sampling indicator turns ON during
measurement.
is displayed rotating during sweep.
Executes source and measure-
ment by one step with trigger
input, and pauses sweep.
The HOLD indicator turns ON.
However, the trigger mode cannot be changed during sweep.