6247G/6247C DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-8
Period
Sets the following period time (Tp).
• Auto sampling period in the DC source mode
• Pulse source period
• Sweep source 1-step period
A.Rng Delay
Sets the wait time (Tar) after changing the range for the measure-
ment auto range.
MEASURE
Sets the measurement-related parameters.
Auto Zero
Switches the measurement auto zero function ON or OFF.
On:
Corrects measurement zero-point drifts approximately
every 10 seconds. The AZ indicator turns ON.
Off:
Turns OFF the auto zero function. The AZ indicator
turns OFF.
Integ Time
Sets the measurement integration time.
Select the integration time from 100
s, 500
s, 1 ms, 5 ms, 10
ms, 1 PLC, 100 ms, 200 ms and S/H.
S/H is the sample hold mode. The integration time is 100
s. The
sample hold mode can be set only when the pulse source mode or
the pulse sweep mode is selected.
The set values are displayed by combining the F, M, S and L indi-
cators as follows:
Table 4-1 Integration Time and Indicator Display
Measure SW
Switches measurement auto ON or OFF.
On:
Executes measurement.
Off:
Does not execute measurement.
Integration time
setting
Indicator
F
M
S
L
S/H
Blink
100
s
Half
500
s
Full
1 ms
Half
5 ms
Full
10 ms
Half
1 PLC
Full
100 ms
Half
200 ms
Full
Blink: Indicator blinks with full brightness.
Half: Indicator turns ON with half brightness.
Full: Indicator turns ON with full brightness.