6247G/6247C DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-12
OPR/SUS IN:
Sets Suspend when the signal level changes from LO to
HI.
Sets Operate when the signal level changes from HI to
LO.
Figure 4-7 OPR/SUS In
OPR/STBY In:
Sets Standby when the signal level changes from LO to
HI.
Sets Operate when the signal level changes from HI to
LO.
Figure 4-8 OPR/STBY In
Cmpl/Sync
Selects the external control signal output function for COM-
PLETE OUT/SYNC OUT.
Meas Front:
Outputs a negative pulse when the measurement starts.
Meas End:
Outputs a negative pulse when the measurement is
completed and the period time ends.
Comp HI:
Outputs a negative pulse when the comparator
calculation result is HI.
Comp GO :
Outputs a negative pulse when the comparator
calculation result is GO.
Comp LO:
Outputs a negative pulse when the comparator
calculation result is LO.
Comp HIorLO:
Outputs a negative pulse when the Comparator
calculation result is HI or LO.
Sync Out:
Outputs a negative pulse when the step starts in the
sweep source mode, or when the pulse source starts in
the pulse source mode.
Sig Width
Selects the control signal pulse width, 10 µs or 100 µs.
OPR/SUS In signal input
Operate status
Operate or
Standby
Suspend
Operate
OPR/STBY In signal input
Operate status
Operate or
Suspend
Standby
Operate