6247G/6247C DC Voltage Current Source/Monitor Operation Manual
3.2 Battery Charge and Discharge Test
3-5
Connecting the DUT
1. Use a 4-wire connection as shown below so that the cables do not cause a voltage
drop.
2. Connect the COMPLETE OUT terminal with the OPERATE IN terminal on the
rear panel by using the BNC-BNC cables A01036.
This is to set the instrument Standby automatically after completing the charge
or discharge.
Figure 3-3 Battery Charge Discharge Test Connection
Charge test
1. Set current source current measurement.
2. Set the charge test parameters.
3. Set the output status to Operate.
The battery starts charging with constant current, and the charge current start
reducing when the battery voltage r 1.45 V.
When the charge current reaches 100 mA or less, the output status becomes
Standby automatically and charging the battery completes.
Battery
+
-
Front
Rear (6247G)