6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.1.2 Screen Display (Annotations)
2-8
FMSL:
Displays the measurement integration time by using the
indicators in combination.
For more information on this combination, refer to
Table 4-1, "Integration Time and Indicator Display."
AUTO: The auto range is set to ON.
4W:
The output sensing is set to 4-wire connection.
2W:
The output sensing is set to 2-wire connection.
MATH: Scaling calculation is ON.
:
Goes ON for every measurement sampling.
:
Indicates the sweep operation status.
Rotates during sweep.
Stops rotation at hold.
Goes OFF when sweep stops.
AZ:
The measurement auto zero function is ON.
NULL: NULL calculation is ON.
ST:
The measurement data memory is ON.
:
Buzzer setting is ON.
(Not in use)
5. Auxiliary indicator for menu
Blinks if a lower hierarchy exists (
key enabled) at parameter
setting on the menu screen.
6. FIT indicator
Source range is set to FIT.
7. HV indicator
High Voltage indicator.
Goes ON when the suspend voltage is +55 V or higher, when the
source value is +55 V or higher in the voltage source function, or
when the limit value is +55 V or higher in the current source func-
tion.
8. Output response indicator
FAST:
The output response is set to Fast.
SLOW: The output response is set to Slow.
9. HL/LL indicator
HL:
The high limit status is detected.
LL:
The low limit status is detected.
10. TpALM indicator
The measurement period becomes longer than the setting Tp. The
following factors are possible:
• Source auto range
• Measurement auto range
• Auto zero
• The measurement time is longer than Tp.
11. Right status indicators
RMT :
Remote control
MA:
Addressed as Lister or Talker via GPIB
SRQ:
SRQ is being transmitted.
MAX, MIN, AVE,
:
MX/MIN calculation is ON.
:
(Not in use)