6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.7 Sweep Measurement
2-45
12. Press
123…
,
1
,
0
and
ENTER
in order.
The stop value is set to 10 V.
13. Press
and
to select
3) Step Value
. Press
to enter the Input/Run
level.
14. Press
123...
,
1
, and
ENTER
in order.
The step value is set to 1 V.
15. Press
and
to select
6) Bias Value
.Verify that the bias value is set to 0 V.
Setting the sweep time
16. Press
and
to select
D) TIME
.
17. Press
and
to select
2) Src Delay
. Press
to enter the Input/Run level.
18. Press
123…
,
1
and
ENTER
in order.
The source delay time is set to 1 ms.
19. Press
and
to select
3) Meas Delay
. Verify that the measurement delay
time is set to 4 ms.
20. Press
twice to select
5) Period
. Press
to enter the Input/Run level.
21. Press
123…
,
1
,
0
,
0
, and
ENTER
in order.
The period time is set to 100 ms.
Setting the measurement data memory
22. Press
,
,
and
in this order to select
F) MEMORY
.
23. Press
to select
1) Store Mode
. Press
to enter the Input/Run level.
24. Rotate
to select
Normal
.
The measurement data memory is set to Normal-ON.