6247G/6247C DC Voltage Current Source/Monitor Operation Manual
A-9
2. Data read time
3. Sweep start data read time (GPIB execution time)
Indicates the time from executing 100-step or 1000-step sweep to completing the data output from
the memory to GPIB with RN1 command, and the collective read-out time.
Conditions: Source range; fixed
Measurement range; fixed, Trigger mode; internal trigger,
Number of measurement digits; 5½ digits, Integration time; 100 µs
Measurement delay; 0.1 ms, Hold time; 1 ms, Source delay; 30 µs
Pulse width; 1 ms
Header; OFF, Block delimiter; EOI (DL2)
Item
Number
of data
GPIB
execution
time
USB
execution
time
Source value data read by query
1
3 ms
20 ms
Measurement buffer memory read after RN1 command
(individual read out)
Condition: Number of measurement digits; 5½ digits,
Header;OFF, Block delimiter; EOI (DL2)
1
3 ms
20 ms
100
131 ms
1.31 s
1000
1.31 s
13.1 s
Measurement buffer memory read after RDT? command
(collective read out)
Condition: Number of measurement digits; 5½ digits,
Header; OFF, Block delimiter; EOI (DL2)
1
3 ms
100
69 ms
1000
626 ms
Number
of steps
Memory mode
Period
Read out
by RN1 command
Collective
read out
100
Normal-ON
10 ms
1135 ms
1070 ms
Burst
2 ms
335 ms
270 ms
1000
Normal-ON
10 ms
11.3 s
10.7 s
Burst
2 ms
3.3 s
2.7 s