4601 I-V Meter Operation Manual
9.7 Source and Measurement Functions
9-8
9.7
Source and Measurement Functions
DC source/measurement:
DC voltage source, DC voltage/current measurement
DC sweep source/measurement:
Source and measurement by linear, 2-slope linear, 3-slope linear,
memory, fixed level
Integration time:
5
s,10
s,25
s,50
s,100
s,250
s,500
s,1 ms,2.5 ms,
5 ms,10 ms,1 PLC,2 PLC,100 ms,200 ms
(PLC: Power Line Cycle 50 Hz: 20 ms 60 Hz: 16.66 ms)
Sweep mode:
Reverse ON (round)/OFF (one-way)
Maximum sweep step:
1999 steps
Maximum measurement points:
2000 points
Maximum sweep memory:
2000 data
Measurement data memory:
4000 data
3
Measurement speed:
50
s/point to 6 s/point (sweep source/measurement mode)
Limit:
HI and LO limit values can be set separately.
(These values cannot be set to the same polarity.)
Trigger:
Auto trigger, external trigger
GPIB
interface:
Compliant with IEEE-488.2-1987
Interface function
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1,
DT1, C0, E2
Connector
Amphenol 24 pin
USB
interface:
USB 2.0 Full-speed
Connector
Type B
External control signal:
TRIGGER IN
OPERATE IN/OUT
COMPLETE OUT, SYNC OUT
Connector
BNC
Contact signal:
Output: 4 bits
Input: 2 bits
Connector
Dsub 15 pin (High-density multicore type)
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......