4601 I-V Meter Operation Manual
5.2.1 DC Source Mode
5-6
Changing the
source value
AUTO
Source value
changing
induces range
changing.
Tp: Period time
Td: Measurement
delay time
Tds:Source delay
time
Tm: Measurement
time
Integration
time + Mea-
surement data
processing
time
Tcn: Operate pro-
cessing time
Trc: Range change
processing
time
HOLD
a. When the sampling is set to AUTO:
• Measurement repeats itself with the specified period time.
• The period time is extended if measurement does not finish within the period time.
b. When the sampling is set to HOLD:
• Measurement starts when the measurement delay time has passed after trigger input.
• Trigger input during measurement is ignored.
c. When the output status is Standby or Suspend:
• Measurement is not executed during Standby or Suspend.
Table 5-2 Operations in DC Source Mode (2 of 2)
Operational
condition
Sampling
Description
Operation
Remarks