4601 I-V Meter Operation Manual
5.3.7 Measurement Data Memory
5-25
5.3.7
Measurement Data Memory
The measurement data memory can store up to 4000 data for each measurement types: ADC1, ADC2 and
ADC3 (in the DC mode, or in the sweep mode with reverse ON). When the memory runs out, memory
storing stops and the MFL bit of the status register is set to ON.
In the DC mode, the memory store function turns ON by selecting
E) MEASURE
from the categories in
the menu and setting
2) Memory Store
to
ON
.
The measurement memory contents are automatically cleared when
2) Memory Store
is switched from
OFF
to
ON
.
In the sweep mode, the memory store function automatically turns ON and cannot be turned OFF.
In the sweep mode, the measurement memory contents are cleared by trigger input. Also, no measured val-
ues are stored in the memory if sweep is paused or stopped.
To clear the memory, execute
3) Memory Clear
.
The following shows the procedure of reading the memory.
Press .
The
MENU
key goes ON.
Example display
Press
four times to show the following display.
Example display
Press .
Example display
Press .
Example display