4601 I-V Meter Operation Manual
8.4.2 Calibration Procedure
8-22
8.4.2.7
Reference Junction Temperature Calibration
The procedure of reference junction temperature calibration is described according to Figure 8-11 Calibra-
tion Procedure (7).
The ADC3 measurement function should be set to T thermocouple measurement.
1. Set the calibration mode to Operate.
2. Select the command depending on either Tc1 or Tc2 function.
Tc1 function:
XRJ1
Tc2 function:
XRJ2
3. Enter the DMM data input mode by XDAT.
4. Set the data (
C) of 0
C standard calibration bath by XD data.
5. Read out and check the measurement data.
6. Move to the next step by XNXT.
7. Set calibration mode to Standby by SBY.