4601 I-V Meter Operation Manual
8.4.1 Calibration Operation
8-14
Calibration procedure (4)
The following figure shows the flow of current measurement calibration.
Connect the 4601 to the standard as shown in Figure 8-1 (b) in the 300
A to 300 mA range or
Figure 8-1 (c) in the 3 A to 10 A range.
Figure 8-8 Calibration Procedure (4)