4601 I-V Meter Operation Manual
6.4 Status Register Structure
6-11
•
Common conditions in which the Status Byte Register is cleared
All cleared when the power is turned ON.
All cleared by *CLS except that MAV is not cleared if data exists in the output buffer.
Not cleared even if read by *STB?.
•
Conditions in which the Service Request Enable Register is cleared
Cleared when the power is turned ON (the PSC flag is set to 1).
Cleared when the *SRE0 command is executed.
The following figure shows the 4601 status register structure.
Figure 6-1 Status Register Structure