4601 I-V Meter Operation Manual
9.6 Temperature Measurement AD590 Measurement Terminal Ad1/Ad2 Terminal
9-7
9.6
Temperature Measurement AD590 Measurement Terminal Ad1/Ad2 Terminal
The accuracy of the temperature sensor AD590 is not included. (Add the AD590 accuracy to the above.)
Maximum allowable input voltage: 4 V peak (terminal-terminal)
310 V peak (terminal-chassis)
The measurement resolution with the integration time of 500
s or less is as follows:
The following full-scale errors (digits) are added in voltage measurement and current measurement with the
integration time of 10 ms or less.
Measurement range
Measurement
resolution
Overall accuracy
Temperature coefficient
(% of r
C)
(ppm of r
C)/
C
AD590
-50.00
C to 150.00
C
0.01
C
0.025+0.1
C
25+0.01
C
Integration time
5
s
10
s
25
s
50
s
100
s
250
s
500
s
Measurement
resolution (digits)
17
9
4
2
2
2
1
Integration time
5
s
10
s 25
s 50
s 100
s 250
s 500
s 1 ms 2.5 ms 5 ms
10 ms
AD590
300
150
60
30
20
10
10
10
5
3
3