4601 I-V Meter Operation Manual
5.3.2 Suspend Function
5-20
NOTE:
The source data during sweeping is generated with the following timing.
1. Standby
Operate
2. Standby
Suspend
3. When Operate is set after changing the sweep parameters in the Suspend status
4. Switching between Suspend and Operate
The sweep data of the 4601 is created basically when the output status is switched from Standby to
Operate or Suspend.
It takes time to create the sweep data because values at each step are calculated and the source values
are saved in the memory.
It takes longer to switch between Standby and Operate because relay ON/OFF operations are re-
quired.
To reduce time and delete these unnecessary relay ON/OFF operation, the output status is switched
between Suspend and Operate.
The source values are not calculated when the output status is switched from Suspend to Operate un-
less the sweep parameters are changed. However, the source values are recalculated when any of the
following items is changed:
Sweep type
Sweep range
First value/second value/third value/last value
Sweep step value
Start address/stop address
Level value
Sample count
Standby status
Sweep source
Operate status
Suspend status
Power ON
Bias output status
Sweep start
Sweep status
Sweep stop
OPERATE
OPERATE
STANDBY
,
OPERATE
STANDBY ,
SUSPEND
SUSPEND
SUSPEND
1
2
3
4
5
6
7
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......