4601 I-V Meter Operation Manual
7.3.8 Type T Thermocouple Temperature Measurement Test
7-4
7.3.8
Type T Thermocouple Temperature Measurement Test
1. Connect the 4601 and a voltage source standard as shown in Figure 8-3 (h).
2. Set the 4601 to the DC source mode, AUTO sampling and integration time of 200 ms.
3. Set ADC3 measurement to the Tc1 or Tc2 function.
4. Set the output status to Operate.
5. Set the source voltage of the standard to 0 V and 10 mV, and verify that the measured values of the
4601 are within the accuracies described in Chapter 9, "SPECIFICATIONS."
The temperature at each voltage is as follows:
7.3.9
External Voltage Measurement Test
1. Connect the 4601 and a voltage source standard as shown in Figure 8-4 (j).
2. Set the 4601 to the DC source mode, AUTO sampling and integration time of 200 ms.
3. Set ADC3 measurement to the external voltage measurement Em and select the 30 mV range.
4. Set the output status to Operate.
5. Generate ZERO and
FS from the standard, and verify that the difference between the voltage source
of the 4601 in the 30 mV to 3 V range and the setting value of the standard is within the accuracy
described in Chapter 9, "SPECIFICATIONS."
Voltage value
0 V
Temperature *
Approx.
C
Voltage value
10 mV
Temperature *
Approx.
C
*When the ambient temperature of the 4601 is 25
C
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......