4601 I-V Meter Operation Manual
5.2.5 Source and Measurement Timings
5-13
•
Period time (Tp)
200
s
Condition)
Source mode: Sweep mode
Sampling: HOLD
•
Measurement delay time (Td) + Integration time (Tit) + AD processing time (Tk) < Period time (Tp)
Condition)
Source mode: DC sweep mode
Integration
time
Tit Time [ms]
Tk time [ms]
Tit+Tk [ms]
50 Hz
60 Hz
50 Hz
60 Hz
5
s
0.005
0.013
0.018
10
s
0.01
0.017
0.027
25
s
0.025
0.027
0.052
50
s
0.05
0.045
0.095
100
s
0.1
0.08
0.18
250
s
0.25
0.04
0.29
500
s
0.5
0.04
0.54
1 ms
1
0.04
1.04
2.5 ms
2.5
0.04
2.54
5 ms
5
0.04
5.04
10 ms
10
0.04
10.04
1 PLC
20
16.666
0.04
20.04
16.706
2 PLC
40
33.333
0.04
40.04
33.373
100 ms
100
0.04
100.04
200 ms
200
0.04
200.04