4601 I-V Meter Operation Manual
6.6.3 Remote Command List
6-33
Remote
Status
*STB?
Query of the status byte register
Response:
ddd
*SRE n
Service request enable register setting (n: 0 to 255)
*6
*SRE?
Response:
ddd
*ESR?
Query of the standard event status register
Response:
ddd
*ESE n
Standard event status enable register setting (n: 0 to 255)
*6
*ESE?
Response:
ddd
MSR?
Query of the measurement event register
Response:
ddddd
MSE n
Measurement event enable register setting (n: 0 to 65535)
(0)
*6
MSE?
Response:
ddddd
QSR?
Query of the questionable event register
Response:
ddddd
QSE n
Questionable event enable register setting (n: 0 to 65535)
(0)
*6
QSE?
Response:
ddddd
*PSC n
When n is other than 0, SRER and SESER are cleared at
power ON.
(1)
*6
When it is set to 0, SRER and SESER are not cleared at power
ON.
*PSC?
Response:
0 or 1 (when n is set to other than 0)
*CLS
Status clear
Operation
complete
*OPC
Sets the LSB of the standard event status register after all
operations are complete.
*OPC?
Response:
1 (after all operations completed)
*WAI
Waits until all operations are complete.(GPIB only)
*9
Calibra-
tion
Calibration data
XINI
Initializes the calibration data area for calibration execution.
(The calibration data in the non-volatile memory is not
affected.)
XWR
Saves the calibration data to the non-volatile memory.
Calibration
execution
XVS
Voltage source function calibration
XILH
Current limit (+) calibration
XILL
Current limit (-) calibration
XVM
Voltage measurement function calibration
XIM
Current measurement function calibration
XIR
Ir measurement function calibration
XEM
External voltage Em function calibration
XT1
Ch1 thermocouple T function calibration
XT2
Ch2 thermocouple T function calibration
XPT1
Ch1 PT function calibration
XPT2
Ch2 PT function calibration
XAD1
Ch1 AD590 sensor function calibration
XAD2
Ch2 AD590 sensor function calibration
XRJ1
Ch1 reference junction temperature calibration
XRJ2
Ch2 reference junction temperature calibration
*6:
Not initialized by the RINI or *RST command.
*9:
Not specified by USB.
Item
Command
Description
Default
Operation
During
calibration
Power
ON
Factory
shipment
DC
OPR/SUS
Sweep
OPR/SUS
Содержание 4601 I-V Meter
Страница 2: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 70: ...4601 I V Meter Operation Manual 3 4 1 Menu Structure 3 18...
Страница 80: ......
Страница 92: ......
Страница 129: ...4601 I V Meter Operation Manual 5 6 1 Synchronous Operation 5 37...
Страница 146: ......
Страница 182: ......
Страница 230: ......