4601 I-V Meter Operation Manual
8.2 Connections
8-2
8.2
Connections
The following figure shows connections for calibrating the 4601 using the digital multimeter 6581, a current
source standard, and a 0
C standard calibration bath.
Figure 8-1 Connections for Calibration (1 of 4)
(a) Voltage source and voltage measurement calibration
6581
4601
Red
Red
Black
Black
A01035, etc.
(b) Current Measurement Calibration (300
A to 300 mA range)
A01035, etc.
Red
Red
4601
Current source standard
(c) Current Measurement Calibration (3 A to 10 A range)
Black
Black
Red
Red
4601
Current source standard
Black
Black
Recommended wire size: 14 AWG
Recommended wire size: 14 AWG