ADASH Ltd.
Adash 4400 – VA4Pro
For further technical and contact information visit www.adash.eu, email: [email protected]
19
Sensors properties
When you connect the sensors to the instrument, you have to tell what kind of sensors you use. The Sensors
button is on most screens in the bottom. You can define the bearing and ISO 10816 group for every AC channel
individually.
Push the Sensors button. In the next menu select channel sensor you want define.
AC1 – AC4
setting of each AC channel properties (sensor, ISO 10816 group, bearing)
All AC Sensors
setting of all AC channels together
All Bearings
setting of one bearing to all AC channels
All ISO
setting of one ISO 10816 group to all AC channels
DC1 – DC4
setting of each DC channel properties
All DC sensors
setting of all DC channels
Tacho
setting of tacho channel properties
Settling Prolongation if you need longer time for sensor settling, use this item
AC sensors
AC (alternate current) sensors are used for signals, e.g. vibrations.
ICP
on, off (selection of required setting accordingly the sensor type)
Sensitivity[mV/unit]
usually 1,10,100, user
Unit
unit selected from the list or user unit
Position
the angle of sensor (see picture bellow). Usually used for proximity sensors.