ADASH Ltd.
Adash 4400 – VA4Pro
For further technical and contact information visit www.adash.eu, email: [email protected]
17
Standard cable wiring
The standard cable, which are the part of the unit, have the sensor signal connected to the pin number 1. The
second sensor wire is connected to the ground (pin 2).
When you use this cable:
In IN1 socket, the signal will be measured on CH1.
In IN2 socket, the signal will be measured on CH2.
In IN3 socket, the signal will be measured on CH3.
In IN4 socket, the signal will be measured on CH4 (ver. 2.0 and higher).
If you want to use the 3-direction sensor, then you should use the IN2 socket ( pins 3,1,5) + ground (pin 2). You
need the special cable for this purpose.