RUN 2 - the TRIAL MASS screen
The trial mass mounting and response measurement is required to take on RUN 2 screen.
Push the Enter Trial button and enter the weight of trial. Confirm by OK. The weight can be negative, it
represents removing of mass ( for example dismounting of old balancing mass).
The recommended weight is displayed only if the Rotor settings parameters are entered.
Push START button for measurement.
Push Edit and all values can be manually entered ( balancing calculator function). Use Tab (in left) for cursor
movement.
After the RUN2 measurement the dynamic factor is displayed. The DFA and DFP ( amplitude and phase)
values are the response values of standardized trial mass ( mm/s
RMS
/ 1kg regardless the selected unit for
balancing). If you will balance the same machine again after some time, then you do not need to measure the
RUN2. Instead of that you only will enter the DFA,DFP values in this screen. Use Enter DF button for that.