TBN580 TRANSMITTER
CALIBRATION
K - 6
Process Calibration
This procedure applies to TBN transmitters configured for either
ANA-
LOG
or
DIGITAL
. Any differences between the two are noted in the
Comments
column.
Key
Display
Comments
Select
YES
.
Select
PROCESS CAL
.
If configured digitally, the
D-TO-A ADJUST
selection
would not appear.
Select
NO
.
If
YES
, a new sensor offset will be calculated and the
cumulative offset change will be reset to 0.00 mV.
Displayed will be the
TEMP COMP TYPE
selected in
configuration.
If
TEMP COMP TYPE
is
MANUAL
or
NONE
(as
selected in configuration), this screen will appear,
otherwise, the next screen,
CHANGE TEMP OF
PROCESS?
will appear. Default is 25
°
C.
The value displayed is the temperature of the pro-
cess (interior of sensor).
NOTE:
If the new temperature value is entered, this
change will affect the calibration of temperature.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
CALIBRATION
→
PROCESS CAL
BENCH CAL pH
BENCH CAL TEMP
D-TO-A ADJUST
E N T E R
INSTALLING A
NEW SENSOR?
→
NO
YES
E N T E R
TEMP COMP TYPE
IN USE:
HIT ENTER
E N T E R
ENTER FIXED
PROCESS TEMP
nnn.n
°
UNITS
E N T E R
CHANGE TEMP
OF PROCESS?
nnn
°
UNITS