TBN581 (ORP/pION) TRANSMITTER
CALIBRATION
L - 7
Select
BENCH CAL.
SAME SENSOR:CALCULATED SENSOR OFFSET
,
CALCULATED EFFICIENCY
, and
CUMULATIVE
OFFSET CHANGE
are recalculated.
NEW SENSOR: CALCULATED SENSOR OFFSET,
CALCULATED EFFICIENCY
are calculated and
CUMULATIVE OFFSET CHANGE
is set to 0.00 mV.
NOTE:
mV SOURCE
selection is covered in
.
Low default value: -1000 mV.
High default value: +1000 mV.
Enter low mV value and press
. Enter high
mV value.
Displays low mV calibration value, and the uncali-
brated mV values determined from the input source.
Displays high mV calibration value, and the uncali-
brated mV values determined from the input source.
Value from last calibration. If new sensor, efficiency
will be 98%.
This value is recalculated during each calibration.
Key
Display
Comments
E N T E R
CALIBRATION
PROCESS CAL
→
BENCH CAL
D-TO-A ADJUST
E N T E R
INPUT SOURCE:
→
SAME SENSOR
NEW SENSOR
mV SOURCE
E N T E R
ENTER LO mV CAL
nn.nn mV
ENTER HI mV CAL
nn.nn mV
ENTER
E N T E R
APPLY nn.nn mV
nn.nn
HIT ENTER
WHEN STABLE
E N T E R
APPLY nn.nn mV
nn.nn
HIT ENTER
WHEN STABLE
E N T E R
CALCULATED
EFFICIENCY
nnn.n %
HIT ENTER
E N T E R
CALCULATED
SENSOR OFFSET
nnn pH
HIT ENTER