TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 10
Key
Display
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select
YES
.
Select
EDIT CAL
.
Select
TEMPERATURE
.
To view values, press
twice.
To change the slope, enter the correct number and
press .
To change the offset, enter the correct number and
press
.
Calibration complete.
C A LI-
B R AT E
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
E N T E R
CALIBRATION
PROCESS CAL
→
EDIT CAL
RESET CAL
E N T E R
EDIT CAL DATA
CONDUCTIVITY
→
TEMPERATURE
E N T E R
TEMPERATURE
PROCESS CAL
SLP: n.nnnn
OFF: n.nnnn
ENTER
ENTER
ENTER
E N T E R
WORKING
CALIBRATION
COMPLETE
E N T E R
ID TAGNAME
READY