![YOKOGAWA ScopeCorder DL350 User Manual Download Page 82](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346082.webp)
4-60
IM DL350-01EN
SENT Data Extraction Conditions (All Sub Channel Setup)
Set the extraction conditions for each data type.
SENT Data Acquisition Destination
SENT data is extracted and acquired in each sub channel. The types of data acquired in sub channels are as
follows.
•
Is Set to OFF
Sub Channel
Data Type
1:FastCH
FastCH
2:FastCH
FastCH
3:FastCH
FastCH
4:S&C
S&C
(Status & Communication)
5:SlowCH
SlowCH
6:SlowCH
SlowCH
7:SlowCH
SlowCH
8:SlowCH
SlowCH
9:SlowCH
SlowCH
10:Error Trigger Error Trigger
11:Error Count
Error Count
• Fast Channel Multiplexing Is Set to ON
Sub Channel
Data Type
1:FastCH
FastCH
2:FastCH
FastCH
3:FastCH
FastCH
4:S&C
S&C
(Status & Communication)
5:SlowCH
SlowCH or FastCH
6:SlowCH
SlowCH or FastCH
7:SlowCH
SlowCH or FastCH
8:SlowCH
SlowCH or FastCH
9:SlowCH
SlowCH or FastCH
10:Error Trigger Error Trigger
11:Error Count
Error Count
Common Settings
Input (Input)
Set the input for data type Fast CH, S&C (Status & Communication), SlowCH, Error Trigger, and Error Count.
ON: The data is monitored.
OFF: The data is not monitored or displayed.
Labels (Label)
You can set the label for data type Fast CH, S&C (Status & Communication), SlowCH, Error Trigger, and Error
Count. You can also set the label for bits 0 to 3 of S&C.
4 Vertical Axis