![YOKOGAWA ScopeCorder DL350 User Manual Download Page 213](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346213.webp)
20-2
IM DL350-01EN
Maximum Number of Waveform Acquisitions
(Maximum number of history waveforms that can be stored in the acquisition memory)
The number of history waveforms that can be stored varies depending on the selected record length as follows:
Record Length
Number of Waveforms
With 4-CH Module
Without 4-CH Module
10 kpoint
1000
1000
25 kpoint
528
1000
50 kpoint
263
528
100 kpoint
105
263
250 kpoint
52
105
500 kpoint
24
52
1 Mpoint
9
24
2.5 Mpoint
4
9
5 Mpoint
1
4
10 Mpoint
1
1
25 Mpoint
1
1
50 Mpoint
1
1
100 Mpoint
−
1
− This record length cannot be set.
Display Range (Start and End Record)
Using record numbers, set the range of history waveforms to display when the display mode is set to All or
Average.
Selectable range: 0 to −(the number of waveform acquisitions − 1)
List of History Waveforms (List)
The history waveform record numbers and the timestamps when the waveforms were acquired are listed.
Clear History (Clear History)
Clears all history waveforms.
You cannot use clear history in the following situations.
• When the instrument is in remote mode.
• When the instrument is printing, when it is executing auto setup, or when it is accessing a storage medium.
• When go/no-go determination is in progress, when action is in progress, or when searching is in progress.
20 History Waveform Display (Scope mode only)