9-2
IM DL350-01EN
Horizontal Cursors (Horizontal) - T-Y waveforms
Two dashed lines (horizontal cursors) appear on the horizontal axis. You can measure the vertical value at the
position of each horizontal cursor and measure the level difference between the horizontal cursors.
Cursor1
Cursor2
Δ
Y
Y1
Y2
Measurement Source Waveform (Trace)
Set the measurement source waveform to one of the waveforms below.
CH1 to CH4
1
, 16chVOLT
2
, 16chTEMP/VOLT
2
, CAN
3
, LIN
3
, SENT
3
, Math1 to Math2
1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2. You cannot
select the input channel of a logic module.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On a model with the /VE option when a CAN bus monitor, CAN & LIN bus, or SENT monitor module is
installed. After you select CAN, LIN, or SENT, select a sub channel. This cannot be selected on a CAN bus
monitor or CAN & LIN bus monitor module if the data type (Value Type) is set to Logic. On a SENT monitor
module, S&C and Error Trigger sub channels cannot be selected.
Moving the Cursors (Cursor1/Cursor2)
Use Cursor1 and Cursor2 to move the cursors.
You can set in the range of –5 to +5 div from the center of the waveform display window. The step resolution is 0.01
div.
Linking Cursors
You can move the cursors without changing the interval between Cursor1 and Cursor2.
Measurement Items (Item Setup)
You can measure the following vertical values at the cursor positions.
Y1
Vertical value at Cursor1
Y2
Vertical value at Cursor2
ΔY
Difference between the vertical values of Cursor1 and Cursor2
9 Cursor Measurement